A Test-A-Pack Model F100-2700-1 Integrity Tester provides a fast and accurate test method for verifying the integrity of non-porous, flexible packages. The system serves medical, pharmaceutical, and food processing applications as a reliable package leak detector.
Features:
The F100-2700-1 Integrity Tester is designed to be used with a package test chamber (sold separately) for testing various packages. It regulates the vacuum source (optional pumps also sold separately) while providing test data on the display screen. There is a check valve located on the tester's chamber port to prevent water contamination during a test.
When using the Integrity Tester in conjunction with a test chamber, the exact location of any package leak can be detected. The test data, time, machine settings and results can be automatically printed out for your records using the optional Test-A-Pack printer. The tester displays readings in inches or millimeters of mercury (gauge or absolute) and can also provide values in feet of altitude up to 50,000 ft.
Specifications:
Vacuum Source
Use Test-A-Pack vacuum pump F136-1115-X or equivalent
Electrical Power
85-264 VAC at
47-63 Hz
Environmental:
Temperature
Operating
Non-operating
40 to 110 deg F
0 to 126 deg F
(4 to 43 deg C)
(-18 to 52 deg C)
Humidity
Up to 90% R.H.
Performance:
Vacuum accuracy
0.2% full scale (50 PSI)
Absolute Pressure (max)
2.4 in Hg
(60 mm Hg)
Absolute Pressure (min)
0.04 in Hg
(1 mm Hg)
Gauge Pressure (max)
29.9 in Hg
(760 mm Hg)
Altitude (max)
50,000 ft
Altitude (min)
Depends on geographic location
Ordering Information:
Part Number
Description
Relevant Dim.
Weight
F100-2700-1
Non-porous Package Integrity Tester
11.4”W x 15.0”L x 6.0”H
12 lbs